HALO® vs CFA

HALO Research Kit Logo

There are now three HALO® Platforms:

  1. HALO®-96 MeC
  2. HALO®-96 SEC, and
  3. HALO®-384 HT

The following table shows the important characteristics between the three HALO® Platforms and how they contrast with the traditional colony-forming assay. This table also provides information that will be important for you to choose a HALO® Platform suited to your specific application. This is also discussed in "Which HALO® Format is Right for You?"

 

Property

Colony-Forming Assay (CAMEO™-4)

HALO®-96 MeC
HALO®-96 SEC
HALO®-384 HT
Type of assay:Differentiation

Proliferation / Cytotoxicity -
Apoptosis

Proliferation / Cytotoxicity -
Apoptosis
Proliferation /
Cytotoxicity - Apoptosis
Validated
No
Yes
Yes
Yes
Type of culture:
Methyl cellulose
Methyl cellulose
Suspension
Suspension
Cell growth:
Clonal
Clonal
Expansion
Expansion
Parameter measured:
Cell colonies
Intracellular ATP
Intracellular ATP Intracellular ATP
Readout:
Manual / Microscopy
Luminescence / Instrument-based
Luminescence / Instrument-based Luminescence / Instrument-based
Subjectivity:
Subjective
Non-subjective
Non-subjective
Non-subjective
Calibration:
Not standardized
External ATP
External ATP
External ATP
Format
35mm Petri dishes
96-well plate
96-well plate
384-well plate
Assay description:
Macro
Mini
Mini
Micro
Volume of assay:
1ml
100µl
100µl
25µl
No. of replicates:
2-3
Min. 4 - unlimited
Min. 4 - unlimitedMin. 4 - unlimited
Cell incubation time:
7-14 days
5-7 days
4-6 days
4-6 days
Turnaround time:
7-14 days
5-7 days
4-6 days
4-6 days
Processing time / Readout time:
None / ~10 min/dish 20 min/plate / <5min/plate
20 min/plate / <5min/plate20 min/plate / <5min/plate
Throughput capability:
Low
Medium
Medium - High
High
Training:
6-12 months
2 days
2 days
2 days